The purpose of this guide is to introduce the LPA3 Portable light extinction particle analyser and help you get started quickly and easily. We map the key areas of the interface and machine layout and take you through an example sample testing.
Full instructions and support are available in the Operator’s Manual at mpfiltri.com
HYDRAULIC CONNECTIONS
HEALTH AND SAFETY
Do not operate, maintain or carry out any procedure before MP Filtri is dedicated to offering comprehensive
reading the Operator’s Manual. It is also important to comply customer support with multimedia guides designed with the current regulations related to occupational accident to help you master the device quickly and easily.
prevention and safety in the workplace.
SUPPORT MATERIALS
You can find all the tools you need at our website: https://www.mpfiltri.com/
Any individual operating the unit shall wear the following
Personal Protective Equipment: Eyewear; Safety shoes;
Gloves; Overalls (or other suitable protective clothing).
A full Operator’s Manual is available
TECHNICAL SPECIFICATION
Particle Sizing
>4, 6, 14, 21, 25, 38, 50, 70 μm
Particle Sizing
±½ ISO code for 4, 6, 14 μm(c)
±1 code for 21, 25, 38, 50, 70 μm(c)
Operating Pressure
2 bar (29 PSI) to 420 bar (6091 PSI)
Reporting Standards
ISO 4406, NAS 1638
AS4059 Rev E, Table 1, 2
AS4059 Rev G, Table 1,2
GBT 14039, GJB 420 B
GOST 17216, ISO 11218
Dimensions
Width: 470 mm (18.5”) - Height: 170 mm (6.7”) - Depth 350mm (13.8") - Weight: 10 kg (22 lbs) - Device only
Data Storage
Approximately 4000 time-stamped tests in the integral memory Full specification available in the Operator’s Manual
LPA3 MAIN SCREEN INTERFACE
LPA3 SETTINGS SCREEN INTERFACE
SAMPLE TESTING
STEP 1
Connect the LPA3 waste outlet quick-release attachment (C) and then connect the waste hose to a
waste receptacle (B)
STEP 2
Connect the hose to the LPA3 high-pressure inlet first (A) then connect to the high pressure test point
on the system to be tested.
STEP 3
The LPA3 is now ready to take samples from pressurised systems.
To begin sampling program the unit via the settings options/screen, purge fluid using the manual flush
option (if required) and then press Start to begin a test.
STEP 4
The results will appear on the main screen. You can press the printer button to get hard copy
results printed out via the onboard printer.
DIMENSIONS